BS&W Analyzer

USE “SEARCH” TAB ON the top right side FOR ALL INSTRUMENTATION RELATED TOPICS

Principal of operation ‘Oscillator Load Pull’  analyzers (BS&W) achieve superior performance by utilizing microwave oscillator load pull. Load pull is the term given to describe the frequency change of an unbuffered oscillator as its output load varies. Circuit components and the external load impedance determine an unbuffered oscillator’s frequency. The permittivity of the materials in the measurement section, through which the microwaves propagate, determine the output load. The measurement section consists of a small solid rod mounted inside a larger diameter pipe, as shown in Figure 1. One end of the rod is connected to an unbuffered oscillator and the other end connects to the center of a welded “shorting” plug. The center rod is covered by a hard plastic sheath to prevent direct contact between the metal rod and conductive water-oil emulsions. Electrically this pipe, rod, and sheath combination is a coaxial transmission line, terminating into a short circuit. The fluids flow through the measurement section via the connections that mount perpendicular to the run section, one at each end. The microwave signal travels the length of the pipe twice; down the pipe from the oscillator, then reflects at the shorting plug and traverses back to the oscillator module

The permittivity of the emulsion changes as the percentage of water in the total fluid changes. The permittivity of the emulsion is comprised of two parts – the dielectric constant and the loss. The relative dielectric constant of oil is 2.2 and of water is about 70. The loss is determined primarily by the salt content of the water. Accurate measurement of the water salinity and proper input to the electronics unit is essential for best accuracy of the Phase Dynamics Full Range WaterHydrocarbon Analyzer. In summary, the permittivity of the oil-water emulsion in the measurement section provides a complex impedance, or load. The load acts directly upon the unbuffered oscillator to force a predictable, repeatable, and precise change in frequency. This frequency is proportional to the water content of the emulsion. Temperature and loss also affect the frequency; both are used for compensation to calculate the correct water content. The microprocessor uses the measured frequency to calculate and update the water content each second.The best make of BSW analyzer is from phase dynamics

 

“In OLP,oscillators frequency changes when its load impedance changes.Here the oscillator operates in microwave frequency and is propageted through the measurement section which is the load and the measurement section is dipped in liquids of permittivities or the dielectric constant of the emulsion.Thus as the impedance changes due to a change in percentage of water in oil,the frequency changes.The frequency and the fluid temperature are measured and gives a value in water content”

To learn more about Instrumentation and Control, purchase your copy now

from amazon

 

23 thoughts on “BS&W Analyzer

  1. Pingback: OIL IN WATER ANALYZER | Kishore Karuppaswamy

  2. Pingback: pH analyzer working principle and Calibration | Kishore Karuppaswamy

  3. Pingback: My Post till now | Kishore Karuppaswamy

  4. Pingback: All my posts till now | Kishore Karuppaswamy

  5. Pingback: A typical Foundation Field bus wiring diagram | Kishore Karuppaswamy

  6. Pingback: Instrumentation related to a motor driven pump | Kishore Karuppaswamy

  7. Pingback: Dampers | Kishore Karuppaswamy

  8. Pingback: Calibration of Temperature transmitter zero trimming | Kishore Karuppaswamy

  9. Pingback: Painting Procedure | Kishore Karuppaswamy

  10. Pingback: Instrumentation Cable design specification | Kishore Karuppaswamy

  11. Pingback: Acceptable accuracy ranges of Instruments | Kishore Karuppaswamy

  12. Pingback: Standard Power supply requirements for Instrumentation devices | Kishore Karuppaswamy

  13. Pingback: General design requirements of Instrumentation part 2 | Kishore Karuppaswamy

  14. Pingback: AK ENCON ENGINEERING SERVICES | Kishore Karuppaswamy

  15. Pingback: Bastard Esterraj Stephen and MC GSN Raju | Kishore Karuppaswamy

  16. Pingback: Boiler | Kishore Karuppaswamy

  17. Pingback: General design requirements of Instrumentation part 5 | Kishore Karuppaswamy

  18. Pingback: All my posts till now | kishore koduvayur

  19. Pingback: AS-i (Actuator sensor-Interface Protocol) | Kishore Karuppaswamy

  20. Pingback: Flow transmitter DP type | Kishore Karuppaswamy

  21. Pingback: Profibus | Kishore Karuppaswamy

  22. Pingback: Calibration of siemens sipart PS2 | Kishore Karuppaswamy

  23. Pingback: Instrumentation engineering | Kishore Karuppaswamy

Leave a comment