The laser level transmitter either uses triangulation, time of flight or confocal chromatic working principle

Triangulation

One method for accurately measuring the distance to targets is through the use of laser triangulation sensors. They are so named because the sensor enclosure, the emitted laser and the reflected laser light form a triangle.

The laser beam is projected from the instrument and is reflected from a target surface to a collection lens. This lens is typically located adjacent to the laser emitter. The camera views the measurement range from an angle that varies from 45 to 65 degrees at the centre of the measurement range, depending on the particular model.

Time of flight

This technique uses the time light takes to travel to a target and back, but the time for a single round trip is not measured directly. Instead, the strength of the laser is rapidly varied to produce a signal that changes over time. The time delay is indirectly measured by comparing the signal from the laser with the delayed signal returning from the target.

Confocal chromatic:

This uses use a white light source to accurately measure the distance to surfaces. The essence of our confocal chromatic imaging principle is the accurate detection of colours from light that is reflected back from target surfaces. The white light is focused onto the target surface by a multi-lens optical system. These lenses disperse the light into monochromatic stages (colours) along the measurement axis. A specific distance to the target is assigned to each colour’s wavelength in a factory calibration. Only the wavelength which is exactly focused on the target is used for the measurement. This light reflected from the target surface is transmitted from the probe, through a confocal aperture and onto a spectrometer which detects and processes the spectral changes and calculates distances. These distance measurements are transmitted at high speed via Ethernet communications protocol

.

Some of the laser LT are below

 

For level transmitter related topics, click on the link below

https://kishorekaruppaswamy.wordpress.com/2017/12/19/magnetostrictive-level-transmitter/

https://kishorekaruppaswamy.wordpress.com/2017/09/28/how-to-line-up-interface-level-transmitter/

https://kishorekaruppaswamy.wordpress.com/2017/08/12/capacitance-type-level-transmitter/

https://kishorekaruppaswamy.wordpress.com/2017/08/11/lrv-and-urv-of-interface-level-transmitter/

https://kishorekaruppaswamy.wordpress.com/2017/05/20/laser-level-transmitters/

https://kishorekaruppaswamy.wordpress.com/2017/05/18/zero-suppression-and-elevation/

https://kishorekaruppaswamy.wordpress.com/2017/01/02/lrv-and-urv-determination-for-d-p-type-level-transmitter/

https://kishorekaruppaswamy.wordpress.com/2016/06/10/level-measurement-using-pressure-gauge/

https://kishorekaruppaswamy.wordpress.com/2016/02/19/calibrationtheory-and-initialization-of-gwr-level-transmitter-rosemount-5300-and-ktek-5100/

https://kishorekaruppaswamy.wordpress.com/2016/02/18/calibration-and-initialization-of-electronic-remote-seal-level-transmitter-rosemount-3051ers/

https://kishorekaruppaswamy.wordpress.com/2016/01/16/calibration-and-initialization-of-wet-leg-impulse-tube-rosemount-3051cd-series-lt/

https://kishorekaruppaswamy.wordpress.com/2016/01/15/calibration-and-initialization-of-rosemount-3051cd-capillary-type-level-transmitter/

https://kishorekaruppaswamy.wordpress.com/2015/02/27/942/

https://kishorekaruppaswamy.wordpress.com/2012/11/16/calibration-procedure-of-displacer-type-level-transmitter/

To learn more about Instrumentation and Control, purchase your copy now

from amazon

 

Advertisements

15 thoughts on “LASER LEVEL TRANSMITTERS

  1. Pingback: Calibration Procedure of displacer type Level transmitter | Kishore Karuppaswamy

  2. Pingback: Magnetostrictive Level Transmitter | Kishore Karuppaswamy

  3. Pingback: Rosemount 3051 series transmitters | Kishore Karuppaswamy

  4. Pingback: How to line up interface Level Transmitter-“Shore Key Line up” | Kishore Karuppaswamy

  5. Pingback: Capacitance type Level Transmitter | Kishore Karuppaswamy

  6. Pingback: LRV and URV of Interface Level Transmitter | Kishore Karuppaswamy

  7. Pingback: ZERO SUPPRESSION AND ELEVATION | Kishore Karuppaswamy

  8. Pingback: LRV and URV determination for D.P. type Level Transmitter | Kishore Karuppaswamy

  9. Pingback: Level Measurement using Pressure Gauge | Kishore Karuppaswamy

  10. Pingback: Calibration,theory and initialization of GWR level transmitter Rosemount 5300 and Ktek 5100 | Kishore Karuppaswamy

  11. Pingback: Calibration and initialization of Wet leg impulse tube Rosemount 3051CD series LT | Kishore Karuppaswamy

  12. Pingback: Calibration and initialization of rosemount 3051CD capillary type level transmitter | Kishore Karuppaswamy

  13. Pingback: Displacer type Level transmitter Fisher FIELDVUE™ DLC3010 Digital Level calibration | Kishore Karuppaswamy

  14. Pingback: My Post till now | Kishore Karuppaswamy

  15. Pingback: All my posts till now | Kishore Karuppaswamy

Leave a Reply

Fill in your details below or click an icon to log in:

WordPress.com Logo

You are commenting using your WordPress.com account. Log Out /  Change )

Google+ photo

You are commenting using your Google+ account. Log Out /  Change )

Twitter picture

You are commenting using your Twitter account. Log Out /  Change )

Facebook photo

You are commenting using your Facebook account. Log Out /  Change )

Connecting to %s